Aurum pre-filled syringes phased introduction of new 10ml CONNECT syringe barrel: Importance of selecting compatible needle-free connectors to minimise the risk of syringe blockage (DSI/2025/02)

moderate

MHRA - Ref: DSI/2025/02

Get recall alerts — sign up free

Hazard

Due to a phased introduction of the new 10ml CONNECT barrel design for the Aurum pre-filled syringes, the range of compatible needle free connectors (NFCs) differs between the established syringe barrel and the new syringe barrel. It is vital that healthcare professionals use compatible NFCs with Aurum pre-filled syringes to prevent delays to delivery of emergency medication.

Description

Update: 22 August 2025 We have issued a minor update to this Device Safety Information to remove the reference to the Naloxone pre-filled syringe as this product, which uses a 2 ml barrel, is not affected by the 10ml CONNECT introduction. All other content remains unchanged. Summary The use of incompatible needle-free connectors (NFCs) with Aurum pre-filled syringes has led to adverse incidents in emergency situations. Attaching an incompatible NFC can block the syringe and prevent delivery of emergency medicine. A new 10ml barrel design for the Aurum pre-filled syringes, CONNECT, will be compatible with a wider range of NFCs to minimise the occurrence of these adverse incidents. For the transition period, both the new syringe and the established syringe will be available for affected products. The CONNECT syringe will be introduced to the market in a phased manner. The 10ml CONNECT barrel for the Adrenaline 1:10,000 pre-filled syringe is scheduled for market introduction in July 2025…

What you should do

Immediate action

Do not stop taking a medicine without advice — speak to your pharmacist or GP. Due to a phased introduction of the new 10ml CONNECT barrel design for the Aurum pre-filled syringes, the range of compatible needle free connectors (NFCs) differs between the established syringe barrel and the new syringe barrel. It is vital that healthcare professionals use compatible NFCs with Aurum pre-filled syringes to prevent delays to delivery of emergency medication.

Return / contact instructions

See the MHRA notice for details.

Product details

  • Brand: MHRA
  • Category: Medicines
  • Product type: Medical device
  • Recall date: 2025-07-01
  • Reference: DSI/2025/02
  • Source: MHRA

Official notice

View the alert on GOV.UK

Data from GOV.UK (Medicines and Healthcare products Regulatory Agency). Ref DSI/2025/02. Contains public sector information licensed under the Open Government Licence v3.0.