BD BodyComm™ version 3.3 software used for BodyGuard infusion pumps, BodyGuard-T and T34 syringe drivers: Important steps to manage transition to Windows 11 (DSI/2025/006)

moderate

MHRA - Ref: DSI/2025/006

Get recall alerts — sign up free

Hazard

The BodyComm software currently uses Windows 10. Microsoft is transitioning to Windows 11 from 14 October 2025 which requires an updated cable to connect the pumps to the BodyComm software. These cables may not be available until after the Windows transition date.

Description

Summary The BodyComm software is used for the servicing and set up of the Bodyguard pumps, Bodyguard-T and T34 syringe drivers. These pumps are used for pain management and other infusion therapies. Windows 10 will no longer by supported by Microsoft from 14th October 2025. An updated cable is therefore required to connect the pumps to BodyComm software, due to the upcoming transition to Windows 11. However, the cables may not be available until after the Windows transition date. This will need effective management to mitigate potential impact to patient treatment. Advice for Clinical Engineers, EBME staff, Medical Device Managers and MDSOs: BD have issued a customer letter advising that all impacted customers should take effective action to allow for continuous use of the BD BodyComm 3.3 software. Identify whether your site uses the BodyComm software for configuration or servicing of BodyGuard pumps, or for BodyGuard-T and T34 syringe drivers (see device details section). If you do h…

What you should do

Immediate action

Do not stop taking a medicine without advice — speak to your pharmacist or GP. The BodyComm software currently uses Windows 10. Microsoft is transitioning to Windows 11 from 14 October 2025 which requires an updated cable to connect the pumps to the BodyComm software. These cables may not be available until after the Windows transition date.

Return / contact instructions

See the MHRA notice for details.

Product details

  • Brand: MHRA
  • Category: Medicines
  • Product type: Medical device
  • Recall date: 2025-10-06
  • Reference: DSI/2025/006
  • Source: MHRA

Official notice

View the alert on GOV.UK

Data from GOV.UK (Medicines and Healthcare products Regulatory Agency). Ref DSI/2025/006. Contains public sector information licensed under the Open Government Licence v3.0.