National Patient Safety Alert: ResMed Astral 100 and 150 Ventilators: Potential for Patient Harm due to Unexpected Interruption of Ventilation Therapy (NatPSA/2026/004/MHRA)

critical

MHRA - Ref: NatPSA/2026/004/MHRA

Get recall alerts — sign up free

Hazard

ResMed has issued a Field Safety Notice regarding Astral 100 and 150 ventilators and associated Printed Circuit Board Assembly spare parts manufactured prior to October 2024. A fault in the device can cause unexpected interruption of ventilation therapy.

Description

Date of Issue: 17 August 2026 Reference No: NatPSA/2026/004/MHRA​ This alert is for action by: all Hospital Trusts, Health Boards and independent healthcare providers delivering NHS, community, home ventilation and private healthcare services. This is a safety critical and complex National Patient Safety Alert. Implementation should be coordinated by an executive leader (or equivalent role in organisations without executive boards). Explanation of identified safety issue ResMed has issued a Field Safety Notice (FSN) regarding Astral 100 and 150 ventilators and associated Printed Circuit Board Assembly (PCBA) spare parts manufactured prior to October 2024. An internal supercapacitor may leak electrolyte, causing damage to specific circuitry of the PCBA and unexpected interruption of ventilation therapy. If the issue occurs while the ventilator is delivering therapy, ventilation therapy stops immediately; a high priority audible alarm activates, and the user interface may display therap…

What you should do

Immediate action

Do not stop taking a medicine without advice — speak to your pharmacist or GP. ResMed has issued a Field Safety Notice regarding Astral 100 and 150 ventilators and associated Printed Circuit Board Assembly spare parts manufactured prior to October 2024. A fault in the device can cause unexpected interruption of ventilation therapy.

Return / contact instructions

See the MHRA notice for details.

Product details

  • Brand: MHRA
  • Category: Medicines
  • Product type: Patient safety
  • Recall date: 2026-08-17
  • Reference: NatPSA/2026/004/MHRA
  • Source: MHRA

Official notice

View the alert on GOV.UK

Data from GOV.UK (Medicines and Healthcare products Regulatory Agency). Ref NatPSA/2026/004/MHRA. Contains public sector information licensed under the Open Government Licence v3.0.