NGPod pH testing device: quarantine and dispose of all devices (DSI/2026/009)

moderate

MHRA - Ref: DSI/2026/009

Get recall alerts — sign up free

Hazard

The NGPod handheld device and NGPod sensor, used to test pH to aid the placement of nasogastric tubes, is being withdrawn from the UK market. The manufacturer has ceased trading in the UK. Users should quarantine all remaining stock and dispose of all NGPod devices.

Description

Summary The NGPod handheld device and NGPod sensor, used to test pH to aid the placement of nasogastric tubes, is being withdrawn from the UK market. The manufacturer has ceased trading in the UK and cannot support continued use of the product. Users should quarantine all remaining stock and dispose of all NGPod devices. While the cessation of trading is not directly related to a current safety concern, a Field Safety Corrective Action (FSCA) was initiated earlier this year, due to two UK adverse incident reports, including one with a fatal outcome. In these instances, the device provided a green result, indicating contact with acidic pH, suggesting that the nasogastric tube was correctly placed when, in fact, it was found in the lung. The Field Safety Notice (FSN) provides further details and should be read in conjunction with this Device Safety Information (DSI). Advice for Healthcare Professionals: stop use of the NGPod handheld device and sensor immediately identify, quarantine an…

What you should do

Immediate action

Do not stop taking a medicine without advice — speak to your pharmacist or GP. The NGPod handheld device and NGPod sensor, used to test pH to aid the placement of nasogastric tubes, is being withdrawn from the UK market. The manufacturer has ceased trading in the UK. Users should quarantine all remaining stock and dispose of all NGPod devices.

Return / contact instructions

See the MHRA notice for details.

Product details

  • Brand: MHRA
  • Category: Medicines
  • Product type: Medical device
  • Recall date: 2026-08-24
  • Reference: DSI/2026/009
  • Source: MHRA

Official notice

View the alert on GOV.UK

Data from GOV.UK (Medicines and Healthcare products Regulatory Agency). Ref DSI/2026/009. Contains public sector information licensed under the Open Government Licence v3.0.