Radiofrequency (RF) Safety Considerations for Wireless Cardiotocography (CTG) Monitoring Systems (DSI/2026/012)
moderate
MHRA - Ref: DSI/2026/012
Get recall alerts — sign up free
Hazard
The MHRA has conducted a review into the use of wireless cardiotocography (CTG) systems that use radiofrequency (RF) telemetry channels. Signal crossover, erroneous CTG traces and signal dropouts have been identified as patient safety risks associated with these devices.
Description
Summary The MHRA has conducted a review into the use of wireless cardiotocography (CTG) systems that use radiofrequency (RF) telemetry channels. Signal crossover, erroneous CTG traces and signal dropouts have been identified as patient safety risks associated with these devices. In rare circumstances, signal crossover may result in the fetal heart rate from one CTG monitor being displayed on another CTG monitor operating on the same RF channel. These risks may result in misinterpretation of CTG traces and inappropriate clinical decisions which could lead to adverse consequences. The risk relates to classic telemetry systems operating within the Industrial, Scientific and Medical (ISM) frequency band where a limited number of channels are available. Actions for Healthcare Professionals including Clinical and Medical equipment management staff: ensure that two or more wireless CTG devices are not operating on the same RF channel unless there is sufficient distance between them to preven…
What you should do
Immediate action
Do not stop taking a medicine without advice — speak to your pharmacist or GP. The MHRA has conducted a review into the use of wireless cardiotocography (CTG) systems that use radiofrequency (RF) telemetry channels. Signal crossover, erroneous CTG traces and signal dropouts have been identified as patient safety risks associated with these devices.
Return / contact instructions
See the MHRA notice for details.
Product details
- Brand: MHRA
- Category: Medicines
- Product type: Medical device
- Recall date: 2026-09-17
- Reference: DSI/2026/012
- Source: MHRA
Official notice
Data from GOV.UK (Medicines and Healthcare products Regulatory Agency). Ref DSI/2026/012. Contains public sector information licensed under the Open Government Licence v3.0.