Trinity Biotech Premier Hb9210™ HbA1c Analyser: Risk of Positive Bias and Updates to Instructions for Use (IFU), including use as a diagnostic aid in diabetes mellitus (DSI/2025/003)

moderate

MHRA - Ref: DSI/2025/003

Get recall alerts — sign up free

Hazard

The MHRA has received reports describing a positive bias in HbA1c results delivered by the Trinity Biotech Premier Hb9210 HbA1c analyser, which has resulted in patients being incorrectly diagnosed as pre-diabetic or diabetic.

Description

Summary The MHRA has received reports describing a positive bias in HbA1c results delivered by the Trinity Biotech Premier Hb9210 HbA1c analyser, which has resulted in patients being incorrectly diagnosed as pre-diabetic or diabetic. The MHRA is working closely with the manufacturer, Trinity Biotech, to resolve this. In addition, Trinity Biotech, has committed to updating the Intended Use statement in the Instructions for Use (IFU), to include the use of the analyser as a diagnostic aide, and to provide further clarity regarding the frequency of preventative maintenance required to ensure the consistent performance of the analyser. Advice for healthcare professionals: the Trinity Biotech Premier Hb9210 HbA1c analyser IFU Intended use statement will be updated to state that results from the analyser can be used as an aid to the diagnosis for diabetes mellitus where HbA1c measurements are taken, ensure that you are familiar with the information presented in the recent Field Safety Notic…

What you should do

Immediate action

Do not stop taking a medicine without advice — speak to your pharmacist or GP. The MHRA has received reports describing a positive bias in HbA1c results delivered by the Trinity Biotech Premier Hb9210 HbA1c analyser, which has resulted in patients being incorrectly diagnosed as pre-diabetic or diabetic.

Return / contact instructions

See the MHRA notice for details.

Product details

  • Brand: MHRA
  • Category: Medicines
  • Product type: Medical device
  • Recall date: 2025-07-22
  • Reference: DSI/2025/003
  • Source: MHRA

Official notice

View the alert on GOV.UK

Data from GOV.UK (Medicines and Healthcare products Regulatory Agency). Ref DSI/2025/003. Contains public sector information licensed under the Open Government Licence v3.0.

All recalls